Hemodialysis Machine ATF1022

Hemodialysis Machine ATF1022: Hemodialysis Machine ATF1022:

touchscreen monitor

* Blood pressure sensor on intake

* Online wireless remote monitoring of dialysis data during dialysis for networked dialysis machines

* Data logging and saving data files/log files for each dialysis session

* Printable graphic dialysis result sheet at the end of hemodialysis process

* Hematocrit module for monitoring blood viscosity/thickness

* Programmable for three types of disinfection operations (Chemical, Thermal and Combined)

* Ability to compute and measure KT/V during hemodialysis process (Online Clearance Monitoring/OCM)

* Adjustable settings for UF and Sodium profiles

* Pressure cuff for monitoring blood pressure

* BTM module for monitoring blood temperature

* Audio and voice alarms

* 45 minutes’ battery backup

 

Dialysis solution temperature 34 to 39 Co
Concentration/ density for dialysis solution 12 to 16 mS/cm
Concentration/ density for bicarbonate solution 2 to 5 mS/cm
Dialysis fluid flow rate 300 to 800 mL/min
Controlling the dewatering process 0.1 to 1.5 L/hr
Intravenous pressure -100 to 400 mmH
Transmembrane pressure (TMP) -50 to 600 mmHg
Blood flow rate 50 to 700 mL/min
Heparin injection rate 0.1 to 10 mL/hr

 

. . / .